DIAL
RAMP

To access the Clinical Menu:
-
Press and hold the RAMP and DIAL for about 5 seconds.
-
You will see "Therapy" displayed on the screen.
MODE
OPTI START
EZ START
AUTO MIN
AUTO MAX
A TRIAL
PRESSURE
-
Auto - treatment at auto-adjusting pressures.
-
C Check - delivers CPAP therapy while automatically adjusting the pressure level to meet patient needs after every 30 hours of therapy use in increments pressure ± 1 cm H2O if needed.
-
CPAP - treatment at fixed pressure.
Starts Auto-CPAP at pressure closer to previous session's 90% pressure to minimize residual events at the beginning of therapy.
-
On
-
Off
Reduces therapy pressure setting for the first few days and gradually increases until prescribed treatment pressure is reach.
-
On
-
Off
Minimum pressure in Auto mode:
-
4.0 - Max Pressure
Maximum pressure in Auto mode:
-
Min Pressure to 20
Will enable the device to deliver Auto CPAP therapy for a selected amount of days.
-
On
-
Off
The treatment pressure in CPAP mode or starting pressure for C Check mode.
-
4 - 20

HEATED HUMIDIFIER

HUMIDIFICATION*
HUMIDIFIER
TUBE TEMPERATURE
SMART RAMP
RAMP TIME
RAMP START
FLEX
Mode of humidification (*available with heated humidifier connected).
-
Adaptive - heating levels automatically adjust based on ambient temperature and humidity levels.
-
Fixed - humidification at constant heat level.
Set desired humidity setting:
-
0-5
Set desired temperature for the heated tube:
-
0-5
When enabled, ramping will occur with auto titrating algorithm during ramp period.
-
On
-
Off
The time period (in minutes) which the pressure increases from a lower pressure to the treatment pressure.
-
0 - 45
Lower pressure which therapy starts at before ramping up to treatment pressure.
-
4 up to treatment pressure
Provides pressure relief at the beginning of therapy:
-
Off
-
C-Flex
-
C-Flex+ - smoother changes in pressures between exhalation and inhalation than C-Flex (only available in CPAP mode).
-
A-Flex - smoother changes in pressures between exhalation and inhalation than C-Flex (only available in Auto mode).
FLEX SETTING
TUBE TYPE
The level of air pressure relief on exhalation:
-
1 - 3
Diameter of tubing that you are using with the device (*disabled with heated tube connected).
-
12
-
15
-
22
TUBE TYPE LOCK
MASK TYPE
MASK TYPE LOCK
CHECK MASK FIT
Enables you to lock tube type setting if you do not want patient to change it:
-
On
-
Off
Resistance control setting for air pressure compensation to support your mask:
-
Off
-
x1 - x5
Enables you to lock make resistance setting if you do not want patient to change it:
-
On
-
Off
You can enable or disable the check mask fit feature:
-
On
-
Off
ADVANCED MENUS
AUTOMATIC ON
AUTOMATIC OFF
LANGUAGE*
You can enable or disable advanced menus for the patient:
-
On
-
Off
You can enable or disable the device to start automatically when patient breathes through the mask:
-
On
-
Off
You can enable or disable the device to stop automatically when patient takes off the mask:
-
On
-
Off
Language displayed on unit. Options will vary by region.
-
English
-
Spanish
CLEAR DEFAULT REMINDERS
Clears default patient reminders:
-
Yes
-
No
RESET DATA
Clears patient data from device and SD card (if installed).
-
Yes
-
No