AirCurve 10 ST-A: Clinical Menu

Resmed.jpg
AirCurve10STA.webp

HOME

DIAL

Available Product Downloads:  ​

PDF.JPG

To access the Clinical Menu:

  • Press and hold the Dial and Home buttons simultaneously for 3 seconds.

Available Modes

  1. CPAP - treatment at fixed pressure

  • Set pressure: 4.0 - 20

  • Mask: Pillows, Full Face, Nasal, Pediatric

2. S - treatment at fixed inspiratory and expiratory pressures

  • IPAP: 4.0 - 30

  • EPAP: 3.0 - 25

  • Ti Max: 0.3 - 4s

  • Ti Min: 0.1 - 4s or Ti Max

  • Rise Time: Min, 150 - 900ms

  • Trigger: Very Low, Low, Med, High, Very High

  • Cycle: Very Low, Low, Med, High, Very High

  • Mask: Pillows, Full Face, Nasal, Pediatric

3. ST - fixed inspiratory and expiratory pressures with backup rate

  • IPAP: 4.0 - 30

  • EPAP: 3.0 - 25

  • iBR: Off or On

  • Target Pt Rate*: 8 - 50 (*available with iBR On)

  • Backup Rate*: 5 - 50 (*available with iBR Off)

  • Ti Max: 0.3 - 4s

  • Ti Min: 0.1 - 4s or Ti Max

  • Rise Time: Min, 150 - 900ms

  • Trigger: Very Low, Low, Med, High, Very High

  • Cycle: Very Low, Low, Med, High, Very High

  • Mask: Pillows, Full Face, Nasal, Pediatric

4. T - fixed rate and inspiratory/expiratory time

  • IPAP: 4.0 - 30

  • EPAP: 3.0 - 25

  • Resp. Rate: 8 - 50 

  • Ti: 0.3 - 0.8s

  • Rise Time: Min, 150 - 500ms

  • Mask: Pillows, Full Face, Nasal, Pediatric

5. iVAPS - maintains minute ventilation by adjusting pressure support and rate

  • Height: 44 - 100inches

  • Target Pt Rate: 8 - 30 

  • Target Va: 1 - 30L/min

  • EPAP: 3 - 25

  • Min PS: 0 - 20

  • Max PS: 0 - 27

  • Ti Max: 0.3 - 4s

  • Ti Min: 0.1 - 4s or Ti Max

  • Rise Time: Min, 150 - 900ms

  • Trigger: Very Low, Low, Med, High, Very High

  • Cycle: Very Low, Low, Med, High, Very High

  • Mask: Pillows, Full Face, Nasal, Pediatric

  • Learn Targets: activate to learn patient's target Va and rate prior to iVAPS mode

  • Ramp Time: Off, 5 - 45min

  • Start Pressure*: 4 - 20 (*enabled with Ramp Time On in CPAP mode)

  • Start EPAP*: 3 - 25 (*enabled with Ramp Time On in S, ST, T, & iVAPS modes)

  • Ramp Down*: Off, On (*available for S, ST, T & iVAPS modes)

  • Climate Ctrl*: Auto, Manual (*enabled with ClimateLineAir heated tubing connected)

  • Tube Temp*: Off, 60 - 86°F (*Climate Ctrl on Manual)

  • Humidity: 0 - 8

Accessories

  • Tube: Slimline, Standard, 3m (*Must be enabled with Essentials Plus.  Disabled with ClimateLineAir connected).  

  • AB Filter: No, Yes

  • Ext. Humidifier: No, Yes

Alarms

  • Low MV: Off, 1 - 10L/min

  • High Leak: Off, On

  • Non-Vent Mask: Off, On

  • Low SpO2: Off, 70 - 95%

  • Apnea Alarm: Off, 10 - 60s

  • Alarm Vol/Test: Low, Med, High

Options

  • Essentials: Plus, On

  • Confirm Stop: Off, On

  • StartStart: Off, On

  • Therapy LED: Off, On

  • Reminders: Mask, Water Tub, Tube & Filter (Off, 1 - 24mths)

Configuration

  • Language: English, Francais, Espanol, Portugues

  • Date: Day, Month, Year

  • Time: Hour, Minutes

  • Desat. Rule: 3%, 4%

  • Press. Units: cmH2O, hPa

  • Temp. units: °C, °F

  • Restore Defaults: No, Yes

  • Erase Data: No, Yes

  • About: Additional information about the unit such as Run Hours and ​internal modem (cellular provider, type and signal).

Therapy 

Accessories, Alarms,Options & Configuration

Comfort

bottom of page