To access the Clinical Menu:
-
At the Home screen, press two fingers anywhere on the screen for about 3 seconds.
-
You will see "Clinical Home" appear at the top left of the screen.
MODE
SET PRESSURE*
-
CPAP - treatment at fixed pressure.
-
AutoSet - treatment at auto-adjusting pressures.
-
AutoSet for Her - treatment at auto-adjusting pressures tailered for female OSA patients with slower pressure increments to reduce arousals.
*For CPAP mode only:
-
4.0 - 20.
PRESSURE RANGE*
*For AutoSet & AutoSet for Her modes only:
-
Min pres: 4 - Max pressure
-
Max pres: Min pressure - 20
AUTOSET RESPONSE*
RAMP TIME
START PRESSURE
EPR
EPR TYPE
HUMIDITY LEVEL
*For AutoSet mode only:
-
Standard
-
Soft
Time period which the pressure increases from a lower starting pressure to the treatment pressure or min pressure:
-
Off
-
Auto
-
5 - 45
Treatment pressure at the start of ramp:
-
4.0 - treatment pressure or min pressure
Expiratory Pressure Relief reduces mask pressure during exhalation for comfort:
-
Off
-
On
-
Ramp Only
-
Full Time
The level the humidifier moistens the air you breathe during treatment.
-
Off
-
1-8

ClimateLineAir 11
Heated tubing

CLIMATE CONTROL*
Controls temperature and humidity levels during therapy (*with ClimateLineAir 11 connected).
-
Auto - automatically adjusts to maintain a constant humidity level of 85%.
-
Manual - you adjust Tube Temperature and Humidity Level separately.
TUBE TEMP.*
The temperature inside the tube (*with ClimateLineAir 11 connected).
-
Off
-
Auto
-
60–86°F
MASK SETTING
TUBE*
B/V FILTER
-
Pillows
-
Full Face
-
Nasal
Inner diameter of tubing used (*disabled with ClimateLineAir 11 connected).
-
SlimLine: 15mm
-
Standard: 19mm
Select whether a bacterial/viral filter is attached
-
Off
-
On
PATIENT VIEW
SMARTSTART
SMARTSTOP
CARE CHECK-IN
REMINDERS
Access level granted to patient at home screen
-
Simple
-
Advanced
If enabled, unit will start automatically when patient breathes into the mask:
-
Off
-
On
If enabled, unit will stop automatically when patient takes off the mask:
-
Off
-
On
Check-in questions during the first month of therapy to provide coaching and support.
-
Off
-
On
Unit will remind patient to replace mask, tube, filter, humidifier:
-
Off
-
1, 3, 6, 9 months, or Yearly.
LANGUAGE
-
English
-
Français
-
Español
-
Português
TIME ZONE
-
GMT -12 to GMT +14
(PST = GMT -8, PDT = GMT -7)
(CST = GMT -6, CDT = GMT -5)
(EST = GMT -5, EDT = GMT -4)
TEMP. UNITS
-
°C
-
°F
RESTORE DEFAULTS
-
Restore settings to factory defaults.
ERASE DATA
-
Erase all data from the unit.
ABOUT
-
Additional information about the unit such as Run Hours and internal modem (cellular provider, type and signal).