top of page
Resmed.jpg
ResmedAir11.jpg

Available Product Downloads:  ​

PDF.JPG

To access the Clinical Menu:

  1. At the Home screen, press two fingers anywhere on the screen for about 3 seconds.

  2. You will see "Clinical Home" appear at the top left of the screen.

MODE

SET PRESSURE*

  • ​​CPAP - treatment at fixed pressure.

  • AutoSet - treatment at auto-adjusting pressures.

  • AutoSet for Her - treatment at auto-adjusting pressures tailered for female OSA patients with slower pressure increments to reduce arousals.

*For CPAP mode only:​

  • 4.0 - 20.

PRESSURE RANGE*

*For AutoSet & AutoSet for Her modes only:

  • Min pres: 4 - Max pressure

  • Max pres: Min pressure - 20

AUTOSET RESPONSE*

RAMP TIME

START PRESSURE

EPR

EPR TYPE

HUMIDITY LEVEL

​*For AutoSet mode only:

  • Standard

  • Soft

Time period which the pressure increases from a lower starting pressure to the treatment pressure or min p​ressure: 

  • Off

  • Auto

  • 5 - 45

Treatment pressure at the start of ramp:

  • 4.0 - treatment pressure or min pressure 

Expiratory Pressure Relief reduces mask pressure during exhalation for comfort:

  • Off

  • On

  • Ramp Only

  • Full Time

The level the humidifier moistens the air you breathe during treatment.        

  • Off 

  • 1-8

climatelineair11.JPG

ClimateLineAir 11

Heated tubing

PDF.JPG

CLIMATE CONTROL*

Controls temperature and humidity levels during therapy (*with ClimateLineAir 11 connected).      

  • Auto - automatically adjusts to maintain a constant humidity level of 85%. 

  • Manual - you adjust Tube Temperature and Humidity Level separately. 

TUBE TEMP.*

The temperature inside the tube (*with ClimateLineAir 11 connected).  

  • Off

  • Auto

  • 60–86°F

MASK SETTING

TUBE*

B/V FILTER

  • Pillows

  • Full Face

  • Nasal

Inner diameter of tubing used (*disabled with ClimateLineAir 11 connected).

  • SlimLine: 15mm

  • Standard: 19mm

Select whether a bacterial/viral filter is attached

  • Off

  • On

PATIENT VIEW

SMARTSTART

SMARTSTOP

CARE CHECK-IN

REMINDERS

Access level granted to patient at home screen

  • Simple

  • Advanced

If enabled, unit will start automatically when patient breathes into the mask:

  • Off

  • On

If enabled, unit will stop automatically when patient takes off the mask:

  • Off

  • On

Check-in questions during the first month of therapy to provide coaching and support.

  • Off

  • On

Unit will remind patient to replace mask, tube, filter, humidifier: 

  • Off

  • 1, 3, 6, 9 months, or Yearly.

LANGUAGE

  • English

  • Français

  • Español

  • Português

TIME ZONE

  • GMT -12 to GMT +14

(PST = GMT -8, PDT = GMT -7)

(CST = GMT -6, CDT = GMT -5)

(EST = GMT -5, EDT = GMT -4) 

TEMP. UNITS

  • °C

  • °F

RESTORE DEFAULTS

  • Restore settings to factory defaults.

ERASE DATA

  • Erase all data from the unit.

ABOUT

  • Additional information about the unit such as Run Hours and ​internal modem (cellular provider, type and signal).

AirSense 11 Autoset: Clinical Home

Therapy

Comfort

Accessories

Options

Configuration