Device Operation
Below is an instructional video on how to use the AirSense 10 AutoSet.
Below are the settings accessible at the My Options screen.
-
TURN the dial to navigate between options.
-
PUSH the dial to select desired option.
RAMP TIME
The time period which the pressure increases from a lower pressure to the prescribed treatment pressure.
-
Off
-
Auto
-
5 - 45min
Recommendation: Start on Auto. If it is ramping up too fast, try a higher ramp time such as 30 - 45 minutes. If it is ramping up too slow, try a lower ramp time such as 5 - 10 minutes or turning ramp Off.
PRESSURE RELIEF*
When turned on, it will allow you to exhale easier (*Must be enabled with Essentials Plus).
-
Off
-
On
Recommendation: Turn On if you have difficulty exhaling while on therapy.
HUMIDITY LEVEL
The level the humidifier moistens the air you breathe during treatment.
-
Off
-
1-8
Recommendation: Start at 4. If your nose/mouth feels dry, increase the humidity level. If you get too much condensation, try using ClimateLineAir heated tubing.

ClimateLineAir
Heated tubing

CLIMATE CTRL*
Controls temperature and humidity levels during therapy (*only available with ClimateLineAir connected).
-
Auto - automatically adjusts to maintain a constant humidity level of 85%.
-
Manual - you adjust Tube Temperature and Humidity Level separately.
Recommendation: Start on Auto. Humidifier output will adjust automatically to maintain a constant humidity level of 85%, which is considered comfortable for most people.
TUBE TEMP*
The temperature inside the tube (*only available with ClimateLineAir connected).
-
Auto
-
60–86°F
-
Off
Recommendation: Start on Auto (with Climate Ctrl on Auto). Tube temperature will be 81°F and Climate Control adjusts the humidifier output to maintain a constant humidity level of 85% while protecting against condensation.
SMARTSTART*
When enabled, unit will turn ON automatically when you breathe with mask on and turn OFF automatically when you remove it (*Must be enabled with Essentials Plus).
-
Off
-
On
MASK
The type of mask you are using with therapy.
-
Pillows
-
Full Face
-
Nasal

TUBE*
Inner diameter of tubing used (*Must be enabled with Essentials Plus. Disabled with ClimateLineAir connected).
-
SlimLine: 15mm
-
Standard: 19mm
RUN MASK FIT
The unit will blow air to check for potential air leaks around your mask.
Recommendation: Run mask fit and adjust mask until you see a green face displayed. If you are unable to get a good mask seal, talk to your provider about how to get a better fit.
RUN WARMUP
The humidifier will rise and hold at a warmer temperature (40°C/104°F) for a total of 30 minutes.
Recommendation: Run warmup before you start therapy if the air is too cold or dry in the beginning of therapy.
AIRPLANE MODE
This disables wireless connectivity and data transmission to AirView and MyAir applications.
-
Off
-
On
Recommendation: Turn On if you need to disable wireless connectivity such as while using your unit on an airplane.
ABOUT*
Here you can view additional information about the unit such as Run Hours and internal modem information (*Must be enabled with Essentials Plus).