top of page

AirSense 10 Autoset: Clinical Menu

Resmed.jpg
AirSense 10 AutoSet .jpg

HOME

DIAL

Available Product Downloads:  ​

PDF.JPG

To access the Clinical Menu:

  • Press and hold the Dial and Home buttons simultaneously for 3 seconds.

MODE

SET PRESSURE

MAX PRESSURE

MIN PRESSURE

MASK
 

  • ​​CPAP - treatment at fixed pressure.

  • AutoSet - treatment at auto-adjusting pressures.

Fixed treatment pressure in CPAP mode:

  • 4.0 - 20

Maximum pressure in AutoSet mode:

  • Min Pressure to 20

Minimum pressure in AutoSet mode:

  • 4.0 - Max Pressure

Type of mask used during treatment:

  • Pillows

  • Full Face

  • Nasal

Comfort

RESPONSE

Rate of pressure increase during therapy in AutoSet mode:

  • Standard

  • Soft

RAMP TIME

Time period which the pressure increases from a lower starting pressure to the treatment pressure or min p​ressure: 

  • Off​

  • 5 - 45

START PRESSURE

Treatment pressure at the start of ramp:

  • 4.0 - treatment pressure or min pressure 

EPR

Expiratory Pressure Relief reduces mask pressure during exhalation for comfort:

  • Off

  • On: Ramp Only

  • On: Full Time

EPR LEVEL

The amount of pressure that is reduced during exhalation:

  • 1

  • 2

  • 3

HUMIDITY

The level the humidifier moistens the air you breathe during treatment:        

  • Off 

  • 1-8

CLIMATE CTRL*

Controls temperature and humidity levels during therapy (*only available with ClimateLineAir heated tubing connected).

  • Auto - automatically adjusts to maintain a constant humidity level of 85%. 

  • Manual - you adjust Tube Temperature and Humidity Level separately.  

TUBE TEMP*

The temperature inside the tube (*only available with ClimateLineAir heated tubing connected):   

  • Auto

  • 60–86°F

  • Off

Accessories

TUBE*

Inner diameter of tubing used (*disabled with ClimateLineAir connected):

  • ​SlimLine: 15mm

  • Standard: 19mm

AB FILTER

Select whether an antibacterial filter is attached:

  • No

  • Yes

Options

ESSENTIALS

SMARTSTART

REMINDERS

Access level granted to patient at home screen:

  • ​Plus: Advanced options and data access

  • On: Basic options and data access

If enabled, unit will start automatically when patient breathes into the mask and stop automatically when mask is removed:

  • ​Off

  • On

Unit will remind patients to replace mask and supplies every 0-24 months: 

  • ​Mask

  • Water Tub

  • Tube

  • Filter

Configuration

LANGUAGE

DATE

TIME

PRESS. UNITS

TEMP. UNITS

RESTORE DEFAULTS

ERASE DATA

ABOUT

  • English

  • ​Francais

  • Espanol

  • Portugues

  • Day / Month / Year

  • Hours : Minutes

  • ​cmH2O

  • hPA

  • °C

  • °F

  • No

  • Yes

  • No

  • Yes

Additional information about the unit such as Run Hours and ​internal modem (cellular provider, type and signal).

Therapy

  • Instagram
  • TikTok
  • YouTube
  • Facebook

© 2023 by CPAPDOCTOR.ORG

bottom of page